SWP-A14-SAIL-methane-sensor-demo
Public demonstration of the SAIL fibre optic methane sensor using balloons filled with 1% CH4 / balance N2 calibration gas (103 L disposable cylinder)
SAFE WORK PROCEDURE

Use this form to document Safe Work Procedures for hazardous activities and processes. The information in your Safe Work Procedure (SWP) should be supported by a risk assessment.

List the Hazards and Risk Controls as per Risk Assessment

Associated risk assessment reference Hazards Risk controls
RA-A14-SAIL-methane-sensor-demo Compressed gas: 103 L disposable steel cylinder at 70 bar with a C10 brass valve Only one cylinder taken to the venue; transported upright in a padded box with the valve closed and regulator removed; kept in shade below 50 degrees C, secured at the back of the table; valve closed between fills; NEVER heated, NEVER refilled
RA-A14-SAIL-methane-sensor-demo Gas mixture: 1% methane in nitrogen (UN1956, Class 2.2 non-flammable, non-toxic); simple asphyxiant only; cannot burn at any dilution in air Open-air venue only; NEVER run inside a small enclosed room with the door shut; SDS No. 303 and the RA provided to the venue and kept at the table; table sign: Calibration gas, 1% methane in nitrogen, non-flammable, Class 2.2
RA-A14-SAIL-methane-sensor-demo Balloon filling: gas jet from the outlet, balloon burst, latex fragments Filled by trained demonstrators only, wearing safety glasses, using a C10 fixed-flow regulator (nominal 1 L/min); filled to no more than 15 cm diameter (1.5 to 2 L); regulator thread checked by hand before opening the valve
RA-A14-SAIL-methane-sensor-demo Public interaction: inhalation of oxygen-free balloon gas, latex allergy, choking hazard for young children, balloon litter Every balloon handed over by a demonstrator, released within arm’s reach and taken straight back; balloons NEVER given away, left on the table or inhaled from; sign: Do not inhale balloon gas; latex-free balloons where practicable or Contains latex notice; children under 12 only with a parent or carer present; fragments binned immediately; balloons counted out and back
RA-A14-SAIL-methane-sensor-demo Electrical equipment outdoors: interrogator, laptop, sample pump, cables Battery power where possible; any mains lead tested and tagged and run through a portable RCD with the connection under cover; cables along the wall or under a cable cover, never across the walkway; stop and cover if rain reaches the table
RA-A14-SAIL-methane-sensor-demo Optical source: fibre-enclosed infrared source (1600 to 1700 nm) inside the interrogator No fibre connector opened while the source is on; fibre run taped down out of public reach; source switched off if any fibre is damaged
RA-A14-SAIL-methane-sensor-demo Public crowd environment: trips, table tipping, heritage venue; weather (heat, rain, wind) Table against the cloister wall, weighted if lightweight; cylinder and pump behind the table; clear 1.5 m walkway; two demonstrators on the stand at all times; cylinder and demonstrators in shade; event marshal directions followed

List Resources Required

Including personal protective clothing, chemicals and equipment needed.

Step by Step Instructions for Undertaking the Task

  1. BEFORE STARTING - CRITICAL SAFETY REQUIREMENTS:
    - You must have completed the SAIL General Local Induction and been briefed on this RA and SWP by the supervisor
    - You must be listed as a rostered demonstrator for this event
    - You must NEVER let anyone inhale from a balloon, and you must NEVER give a balloon away
    - You must NEVER leave the cylinder with the regulator fitted and the valve open while unattended
    - You must NEVER expose the cylinder to heat, direct sun or a parked vehicle
    - You must NEVER open a fibre connector while the interrogator is switched on
  2. PRE-EVENT PREPARATION (in the laboratory, the day before):
  3. Check the cylinder:
    - Confirm the label reads 1% CH4 / Bal. N2, part 103DS1VCH4N, SDS 303, Class 2.2
    - Check the cylinder body for dents, corrosion or damage to the C10 valve outlet thread
    - Check the shelf-life date on the label has not passed
    - If ANY check fails: DO NOT USE, tag the cylinder and report to the supervisor
  4. Test-fit the regulator:
    - With the cylinder upright, screw the C10 regulator onto the valve outlet by hand until fully seated; the thread must engage smoothly without force
    - Open the cylinder valve slowly (anticlockwise, one full turn); the regulator gauge should read approximately 70 bar (lower if partly used)
    - Close the valve and listen for leaks at the joint; if hissing is heard, close the valve, remove and re-seat the regulator
    - Operate the flow lever and confirm gas flows from the nozzle, then close it
    - Close the cylinder valve, bleed the regulator until the gauge reads zero, and remove the regulator for transport
  5. Trial-fill two balloons to approximately 15 cm diameter (about 1.5 to 2 minutes each at 1 L/min). Tie off and confirm they hold gas for at least an hour. Note how many seconds gives the right size
  6. Check the sensor apparatus:
    - Charge the laptop and any battery packs
    - Confirm the interrogator label laser class and that all fibre connectors are mated and the fibre run is secured
    - Run the sensor software, release a trial balloon at the intake and confirm the reading responds
    - Confirm any mains equipment carries a current test and tag label
  7. Pack the kit against the resources list, including signs, SDS, RA, first aid kit and balloon count sheet
  8. Brief all demonstrators (30 minutes): hazards, the balloon hand-over rule, the no-inhalation rule, emergency contacts and the location of the event first aid post
  9. TRANSPORT TO THE VENUE:
  10. Close the cylinder valve and confirm the regulator is removed
  11. Place the cylinder upright in the padded transport box with the valve protected
  12. Carry the box by hand or on a trolley. Do not place it in a backpack, do not throw or roll it, and do not leave it in a vehicle in the sun
  13. On arrival, place the box in a shaded position behind the table
  14. SETTING UP THE DISPLAY (MANDATORY - DO NOT SKIP):
  15. Sign in with the event coordinator and record the start time and demonstrator names on the balloon count sheet
  16. Position the table against the cloister wall (or under the marquee on Eastern Avenue) so the public approach from one side only. Weight the table if it is lightweight. Keep a clear 1.5 m walkway past the display
  17. Set up the sensor:
    - Place the interrogator, laptop and pump at the back of the table with the fibre run taped down and out of reach
    - Route the sample intake tube to the front edge of the table where the public will release balloons
    - Use battery power. If mains is unavoidable, plug the single tested and tagged lead into the portable RCD, press the RCD TEST button and confirm it trips, reset it, and cover the connection. Run the cable along the wall under a cable cover
    - Switch on the interrogator and laptop and confirm a stable baseline reading
  18. Set up the cylinder:
    - Stand the cylinder upright in its box behind the table, in shade, where it cannot be knocked over
    - Touch-check the cylinder is not warm
    - Fit the C10 regulator by hand until fully seated
    - Open the cylinder valve slowly one full turn, check the gauge, then close the valve again. Leave the valve CLOSED until filling
  19. Display the signs (Non-flammable calibration gas; Do not inhale balloon gas; Contains latex if applicable) and place the SDS and RA in a folder on the table
  20. Confirm emergency arrangements: each demonstrator knows the location of the event first aid post, the nearest exit and assembly point, and has the emergency contact card
  21. Final check before opening:
    - Cylinder valve closed, cylinder secured and shaded
    - Cables clear of the walkway
    - First aid kit and eye rinse at the table
    - Lidded bin at the table
    - Two demonstrators present
    - If ANY check fails: DO NOT OPEN THE DISPLAY; resolve the issue or contact the supervisor
  22. FILLING BALLOONS (demonstrators only):
  23. Put on safety glasses and stand at the back of the table facing away from the public
  24. Stretch a balloon over the regulator nozzle and hold the neck firmly
  25. Open the cylinder valve one full turn
  26. Operate the flow lever for the pre-timed number of seconds to reach approximately 15 cm diameter. NEVER fill beyond 15 cm; a small balloon is enough for the sensor and will not burst
  27. Release the lever, pinch the neck, remove the balloon and tie it off
  28. Repeat for a batch of 10 to 15 balloons, then CLOSE THE CYLINDER VALVE
  29. Inspect each balloon; discard any that are weak, perished or over-stretched into the lidded bin
  30. Store filled balloons in a lidded tub or tied to a weighted string behind the table (the gas does not float, but wind can move them). Record the number filled on the balloon count sheet
  31. Close the cylinder valve at any time the table is unattended or filling stops for more than a few minutes
  32. RUNNING THE DEMONSTRATION (public interaction):
  33. ATTENDANCE REQUIREMENT: two demonstrators MUST be on the stand at all times during opening hours. One runs the sensor and hands balloons over; the other manages the queue and fills balloons
  34. For each participant:
    - Hand over one balloon and say: squeeze the neck and let the gas out near this tube, like a fart; please do not breathe it in, there is no oxygen in it
    - Stay within arm’s reach while they release the gas at the intake tube
    - Point out the sensor reading rising on the laptop screen and explain the science
    - Take the balloon straight back. Balloons are NEVER given away, left on the table, or handed to a second person unsupervised
  35. For children under 12: only with a parent or carer present. For young children, the demonstrator holds the balloon and lets the child squeeze it
  36. If a balloon bursts: pick up every fragment immediately and bin it. Do not let children pick up fragments
  37. If anyone says they have a latex allergy: offer the demonstrator-operated release, or use a latex-free balloon
  38. If anyone inhales from a balloon: take it away, sit them in fresh air away from the crowd and follow the emergency procedure
  39. Periodic checks every 30 minutes:
    - Cylinder valve closed and cylinder not warm
    - Balloon count (out and back) matches
    - Cables still covered and walkway clear
    - No balloon fragments on the ground
    - Sensor reading has returned to baseline between participants (if not, move the intake tube into cleaner air for a minute)
  40. If it rains on the table, or wind makes the balloons or equipment unmanageable: close the valve, switch off and cover the equipment and pause the demonstration
  41. PACK-DOWN:
  42. Close the cylinder valve fully (clockwise)
  43. Bleed the regulator by operating the flow lever until the gauge reads zero, then unscrew the regulator
  44. Place the cylinder upright in its padded box
  45. Deflate all remaining balloons (cut the neck) and put all balloon material in the lidded bin. Reconcile the balloon count sheet; leave NO balloon material at the venue
  46. Switch off the interrogator, pump and laptop. Unplug and coil any mains lead and the RCD. Do not open fibre connectors at the venue
  47. Remove all signs, tape and cable covers, leaving the heritage surfaces as found
  48. Sign out with the event coordinator; note any incidents or near misses on the count sheet and report them to the supervisor
  49. Transport the cylinder and equipment back to A31 P2012 in the padded box, and return the cylinder to the gas storage cabinet with the valve closed, regulator removed and label facing out

Emergency Shutdown Procedures

Emergency Procedures for Fires, Spills or Exposure to Hazardous Substances

Clean Up and Waste Disposal Requirements

References Used in the Development of This SWP

E.g. codes of practice.

Competency Required

Qualifications, certificates, licensing, training — e.g. course or instruction.

Staff Approved to Assess Competence for This SWP

Chris Betters (Senior Research Fellow, SAIL)

SWP Sign Off Sheet

SWP name and version: Public demonstration of the SAIL fibre optic methane sensor using balloons filled with 1% CH4 / balance N2 calibration gas (103 L disposable cylinder) (Version 1.0)

In signing this section the assessor agrees that the following persons are competent in following this SWP.

Competency sign-off is recorded on the printed copy of this SWP (use the Word/PDF download) or in the laboratory training register.

Name Signature Date Competent Name of Assessor/Authoriser Assessor/Authoriser Signature
         
         
         
         
         

End of Safe Work Procedure